1) New AAPS Recommendations on Analytical Instrument Qualification
In July 2003 the Gold Sheet published a Summary Report on AAPS
Instrument Qualification Workshop (March 2003). This workshop was
co-sponsored by ISPE and the International Pharmaceutical Association
(FIP). The aim of this meeting, including FDA representatives, was to
establish common terminology for Analytical Instrument Qualification
(AIQ).
It was considered appropriate that the DQ, IQ, OQ, PQ terms for
qualification – although these terms were found confusing for AIQ –
should be continued and that the AIQ process should be defined around
them, since these terms have already found their place in the validation
(qualification) terminology.
Key requirements for IQ, OQ and PQ were defined. For instance the term Operational
Qualification (OQ) is defined and explained as:
- On-site testing and documentation providing verification that an
analytical instrument is functioning as the vendor intended
- OQ may be performed at some regular interval thereafter
- OQ is also performed as required by the change control of the
instrument, and after major repairs that might affect critical factors of
instrument operation
- OQ tests are not method specific
- Absorbance detector wavelength accuracy and absorbance linearity,
injector precision, pump flow rate accuracy and proper gradient operation
are some important tests for HPLC components
- Holistic tests are acceptable in lieu of modular testing
Performance Qualification is defined as on-site testing that
verifies consistent and proper operation of the analytical instrument or
system, according to a specification appropriate for its routine use.
These tests check the maintenance and calibration of the instrument
System. Suitability checks can supplement but not replace the functional
checks for PQ!
Furthermore, software validation was probably the most controversial
issue handled at this AAPS workshop. A white paper will be published based
on the workshop discussions.
Equipment Qualification (including the a.m. AIQ proposals) and Software
Validation are two of the main topics of the Conference of the
European Compliance Academy Successful HPLC Management in a
GMP-/FDA-Regulated Environment. This conference takes place on 31
March and 1 April 2004 in Barcelona, Spain followed by a
Post-Conference Workshop on Validation of Chromatography Data Systems. Click
here to find the complete programme.
2) HPLC in Human Drug cGMP Notes
In earlier issues of FDA's Human Drug CGMP Notes there
were two questions dealing with HPLC, which might be interesting although
they are not brand-new:
The first question was in March 1996:
"Does FDA have a CGMP policy on use of recycled solvents for HPLC
columns?"
The answer was: "The agency has no specific policy on recycled
HPLC solvents…. Therefore it would be acceptable to use recycled
solvents which do not interface with analytical results or equipment
operation."
The second question was in September 1997:
"When analyzing a sample by HPLC to determine variance within a
batch, such as content uniformity or dissolution, and the result for one
tablet is out of specifications (OOS), if there is clearly an HPLC
malfunction, does the whole test need to be re-run, or only the one
solution?"
Russ Rutledge (FDA) answered to this question:
"……if the analytical instrument was clearly shown to give
unreliable results by the laboratory investigation, then the test results
must all be invalidated and the entire test run again, using the original
sample solutions if possible. The new results are substituted for the
invalidated results, and only these are used in the batch evaluation. To
merely re-run the one sample solution is inadequate because the rejection
of data is based on instrument malfunction. Thus, all results from this
test should be considered unreliable."
One lecture at the international HPLC Conference Successful HPLC
Management in a GMP-/FDA-Regulated Environment in Barcelona will cover
Out of Specification HPLC Results. Click here for the complete
programme.
3) HPLC in FDA Warning Letters
FDA Inspections often cover HPLC details as you can see from the below
listed FDA Warning Letter citations:
- Procedure for HPLC Calibration Method 013 has no predetermined
acceptance criteria for the auto sampler calibration. The auto sampler
calibration does not demonstrate that the instrument is capable of
accurately assessing linearity
- The calibration procedure for HPLC Systems is inadequate in that it
did not include integrator and detector's linearity, injector's
reproducibility, and accuracy of temperature settings for column
heater and detector
- Data from two of the test runs were voided due to HPLC systems
problems, however a review of the chromatograms showed that system
suitability was maintained throughout the run
- No data demonstrating suitability for use could be provided for your
firm's HPLC system
- HPLC chromatograms were found to have poor separation between active
ingredient peak and unknown peak, as well as widely different
retention times for the active peak between different chromatographic
runs
- HPLC chromatographic data do not always document system suitability
requirements